Echo-planar imaging (EPI) sequences were developed for a 9.4 Tesla vertical standard bore (∼54 mm) superconducting magnet using an unshielded gradient coil optimized for live mice imaging and a data correction technique with reference scans. Because EPI requires fast switching of intense magnetic field gradients, eddy currents were induced in the surrounding metallic materials, e.g., the room temperature bore, and this produced serious artifacts on the EPI images. We solved the problem using an unshielded gradient coil set of proper size (outer diameter = 39 mm, inner diameter = 32 mm) with time control of the current rise and reference scans. The obtained EPI images of a phantom and a plant sample were almost artifact-free and demonstrated the promise of our approach.
机译:使用针对活小鼠成像优化的非屏蔽梯度线圈和参考扫描数据校正技术,针对9.4特斯拉垂直标准孔(〜54 mm)超导磁体开发了回波平面成像(EPI)序列。由于EPI需要快速切换强磁场梯度,因此在周围的金属材料(例如室温孔)中会感应出涡流,这会在EPI图像上产生严重的伪影。我们使用适当大小的非屏蔽梯度线圈组(外径= 39 mm,内径= 32 mm)并通过电流上升和参考扫描的时间控制来解决该问题。获得的模型和植物样品的EPI图像几乎没有伪影,证明了我们的方法的希望。