>Background: Kirschner wires (K-wires) are commonly used in orthopedic surgery. However, the loosening of the pins can lead to delayed or improper healing or infection. Wire loosening can occur by thermal necrosis that occurs due to heat produced during wire insertion. Although the parameters that affect temperature rise in cortical bone during wire insertion and drilling have been studied, the effect of drilling mode (oscillation versus forward) is unknown. The purpose of this study was to compare the temperature changes occurring in cortical bone during wire insertions by oscillating and forward drills. Our hypothesis is that oscillation drilling would produce less heat compared with forward drilling in K-wire insertion with 2 commonly used wire diameters. >Methods: We drilled K-wires in a pig metacarpal model and measured the temperature rise between forward and oscillation drilling modes using diamond-tipped 0.062- and 0.045-inch-diameter K-wires. There were 20 holes drilled for each group (n = 20). >Results: The average temperature rise using the 0.062-inch K-wire under forward and oscillation insertion was 14.0 ± 5.5°C and 8.8 ± 2.6°C, respectively. For the 0.045-inch K-wire, under forward and oscillation insertion, the average temperature rise was 11.4 ± 2.6°C and 7.1 ± 1.9°C, respectively. The effects of the drilling mode and wire diameter on temperature rise were significant (P < .05). >Conclusions: In conclusion, the oscillation of K-wires during insertion causes a lower temperature rise when compared with forward drilling.
机译:>背景:克氏针(K线)通常用于整形外科。但是,销子松动会导致延迟或不适当地的愈合或感染。由于电线插入过程中产生的热量而导致的热坏死会导致电线松动。尽管已经研究了在导线插入和钻孔过程中影响皮质骨温度升高的参数,但是未知钻孔模式(振荡与正向)的影响。这项研究的目的是通过摆动和向前钻比较在电线插入期间皮质骨中发生的温度变化。我们的假设是,与2种常用线径的K线插入中的正向钻孔相比,振荡钻孔产生的热量更少。 >方法:我们在猪掌骨模型中钻出了K线,并使用直径为0.062英寸和0.045英寸的金刚石尖端的K线测量了正向和振荡钻削模式之间的温度升高。每组钻了20个孔(n = 20)。 >结果:使用0.062英寸K线在正向插入和振荡插入下的平均温升分别为14.0±5.5°C和8.8±2.6°C。对于0.045英寸K线,在正向插入和振荡插入的情况下,平均温升分别为11.4±2.6°C和7.1±1.9°C。钻孔方式和金属丝直径对温度升高的影响显着(P <.05)。 >结论:总而言之,与正向钻孔相比,插入过程中K线的振荡引起较低的温度升高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号