摘要

The correct evaluation of the healing process is important to define proper times of fixator dynamization and removal, avoiding refractures. Unfortunately, a quantitative healing assessment is not yet available in clinical practice. The aim of the paper is to prove the feasibility of the mechanical vibration method to assess bone healing in fractures treated with external fixation, in in vivo conditions. The case study was a patient with a tibial fracture treated with a monoaxial fixator. The healing process was monitored for three months through a series of five impact tests. The pins screwed into the bone were used both to excite and measure vibrations. Fracture healing was quantitatively assessed by estimating the resonant frequencies of the leg. The first frequency increased of about 4% per week during the observation period. After the hard callus formation (13 week), also other frequencies increased within the range 1%–6% per week. X-ray observations confirmed the healing progress and proved the method potentiality. In addition, the vibratory response of the leg after fixator removal was evaluated and resulted characterized by five modes in the bandwidth 0–1000 Hz. The results suggest that the vibratory response of a fractured bone treated with external fixation can be a promising indicator for quantitative healing monitoring. The mechanical vibration method could be helpful for reducing X-ray exposure of patients and could be performed more frequently, as desirable for obtaining more attentive monitoring.
机译:对愈合过程的正确评估对于定义正确的固定器动态化和去除时间(避免骨折)很重要。不幸的是,临床实践中尚无定量的愈合评估。本文的目的是证明机械振动方法在体内条件下评估外固定治疗骨折的骨愈合的可行性。案例研究是使用单轴固定器治疗胫骨骨折的患者。通过一系列的五个冲击测试,对愈合过程进行了三个月的监控。拧入骨头的销钉既用于激发振动,又用于测量振动。通过估计腿的共振频率定量评估骨折愈合。在观察期间,第一频率每周增加约4%。硬call形成后(13周),其他频率也每周增加1%–6%。 X射线观察证实了治疗进展,证明了该方法的潜力。另外,评估了去除固定器后腿的振动响应,并以带宽0–1000 Hz中的五种模式进行了表征。结果表明,经外固定治疗的骨折骨的振动反应可能是定量愈合监测的有希望的指标。机械振动方法可能有助于减少患者的X射线暴露,并且可以更频繁地执行,这对于获得更细心的监视是合乎需要的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号