Cause of the power-on reset phenomenon other than electric magnetic interference in a patient with a pacemaker

机译:带起搏器的患者中除电磁干扰以外还有上电复位现象的原因

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A 67-year old male with a dual-chamber pacemaker visited for a regular check-up. An unfamiliar message emerged on the display just after placing the programmer wand. We could recognize that the pacemaker had already been in the safe back-up mode of DDI, and the programmer prompted a re-initialization request. We are so surprised because there was no indication of device malfunction the day before in daily monitoring and a 12-lead electrocardiogram revealed normally working in the DDD mode just before checking the device. The pacemaker was immediately re-programmed to the former setting. This phenomenon has not recurred for 12 months.
机译:一名67岁的男性与一个双室起搏器进行了定期检查。放置编程器棒后,显示屏上出现一条陌生消息。我们可以认识到起搏器已经处于DDI的安全备份模式下,并且程序员提示了重新初始化请求。我们感到非常惊讶,因为在日常监控中前一天没有任何设备故障的迹象,并且在检查设备之前显示了12导联心电图在DDD模式下正常工作。起搏器立即被重新编程为以前的设置。这种现象已经十二个月没有复发了。