Whereas buckling can cause type III endoleaks, long-term twisting of a stent-graft was investigated here as a mechanism leading to type V endoleak or endotension. Two experimental device designs supported with Z-stents having strut angles of 35° or 45° were compared to a ringed control under accelerated twisting. Damage to each device was assessed and compared after different durations of twisting, with focus on damage that may allow leakage. Stent-grafts with 35° Z-stents had the most severe distortion and damage to the graft fabric. The 45° Z-stents caused less fabric damage. However, consistent stretching was still seen around the holes for sutures, which attach the stents to the graft fabric. Larger holes may become channels for fluid percolation through the wall. The ringed stent-graft had the least damage observed. Stent apexes with sharp angles appear to be responsible for major damage to the fabrics. Device manufacturers should consider stent apex angle when designing stent-grafts, and ensure their devices are resistant to twisting.
机译:屈曲可引起III型内漏,而在这里研究了支架移植物的长期扭曲是导致V型内漏或内张力的一种机制。将两种实验装置设计与支杆角度为35°或45°的Z形支架支撑在一起,并与在加速扭曲下的环形控件进行了比较。在扭曲了不同的时间后,评估并比较了每个设备的损坏情况,重点放在可能会泄漏的损坏上。具有35°Z形支架的支架移植物最严重地变形和损坏了移植物织物。 45°Z型支架对织物的损害较小。然而,在缝合线的孔周围仍然看到一致的拉伸,这些缝合线将支架附接到移植物织物上。较大的孔可能会成为流体渗入壁的通道。环状支架移植物的损伤最小。尖角的支架顶端似乎是造成织物严重损坏的原因。设备制造商在设计支架植入物时应考虑支架的顶角,并确保其设备抗扭曲。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号