Performance of exhaust-protective (class I) biological safety cabinets.

机译:排气防护(I类)生物安全柜的性能。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Nineteen open-fronted (class I) safety cabinets were subjected to spore containment and airflow tests, which suggested that a flow of 0.75 m/s was the minimum required for safe operation; further tests on three of the cabinets were repeated at different air speeds and confirmed this. The airflow is required to overcome the effect of laboratory or external air currents. Contamination of surfaces (including the operator's hands) by aerosols liberated inside the cabinets was investigated and found to depend more on turbulence within the cabinet (as from a restricted front opening) than on the number of air changes. The findings suggest that the product protection provided by a well-designed class I cabinet might equal that of a class II (laminar flow) unit.
机译:对19个敞开式(I类)安全柜进行了孢子密闭性和气流测试,这表明安全操作所需的最低流量为0.75 m / s。以不同的风速对三个机柜进行了进一步测试,并证实了这一点。需要气流以克服实验室或外部气流的影响。调查了机柜内部释放的气溶胶对表面(包括操作员的手)的污染,发现该问题更多地取决于机柜内部的湍流(例如从前部受限的开口),而不是取决于换气次数。研究结果表明,精心设计的I类机柜提供的产品保护可能等同于II类(层流)装置的产品保护。