首页> 中文期刊> 《等离子体科学和技术:英文版》 >Influence of Working Pressure on Ion Sensitive Probe Measurement in Microwave ECR Plasmas

Influence of Working Pressure on Ion Sensitive Probe Measurement in Microwave ECR Plasmas

         

摘要

In order to precisely measure the ion parameters in a microwave electron cyclotron resonance plasma using an ion sensitive probe,the dependences of the current-voltage(I-V)characteristics on the shielding height(h)and the potential difference between inner and outer electrodes(V_B)have been investigated at different working pressures of 0.03 Pa and 0.8 Pa.Results show that the I-V curves at higher pressure are more sensitive to the variation of h than those at lower pressure.The influence of V_B on ion temperature(T_i)measurement becomes more prominent when the pressure is increased from 0.03 Pa to 0.8 Pa.Under both pressures,the optimized h is obtained at the condition where the current reaches zero in the positive voltage region with a suitable V_B of-1.5 V because of effective shielding of the electron E×B drift.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号