Energy response correction for an electronic personal dosimeter using the channel ratio method

         

摘要

An electronic personal dosimeter mainly uses a Si-PIN photodiode as X-and gamma-ray detectors.The photon energy response of this instrument is inconsistent in the case of no correction,which seriously affects the accurate monitoring of personal dose equivalent H_p(10)parameters for radiation workers.For this reason,in this paper we propose a method of combining composite screen detection technology,multichannel measurement technology,and the channel ratio method to achieve accurate measurement of the personal dose equivalent parameters.According to China National Standard GB/T 13161-2003 and National Verification Regulation JJG 1009-2006,the instrument was tested in the energy range between 48 keV and 1.25 MeV.The experimental results showed that the difference of energy response to ^(137)C_S corrected by the new method was almost constant within ±6.0%,which fulfilled the ±30% requirement of GB/T 13161-2003 and JJG1009-2006.Meanwhile,the method proposed obtained energy information regarding the radiation field.