Volume conduction energy transfer for implantable devices

机译:植入式设备的体积传导能量转移

获取原文
获取原文并翻译 | 示例
       

摘要

A common model of power supply for implantable devices was established to study factors affecting volume conduction energy transfer.Electromagnetic and equivalent circuit models were constructed to study the effect of separation between the source electrode pairs on volume conduction energy transfer.In addition,the parameters of external signal including waveform,amplitude and frequency were analyzed.As the current amplitude did not lead to tissue injury and the current frequency did not cause nerve excitability,the recommended separation between the source electrodes was 3 cm,the proposed waveform of signal source was sinusoidal wave and the optimal frequency was 200 KHz.In agar experiment and swine skin experiment,the current transfer efficiencies were28.13%and 20.65%,respectively,and the energy transfer efficiencies were 9.86%and 6.90%,respectively.In conclusion,we can achieve optimal efficiency of energy transfer by appropriately setting the separation between the source electrode parameters of the signal source.