首页> 中文期刊> 《工程(英文)(1947-3931)》 >Enzyme Biosensor for Detection of Organophosphate Pesticide Residues Base on iScreen Printed Carbon Electrode/i(SPCE)-Bovine Serum Albumin (BSA)

Enzyme Biosensor for Detection of Organophosphate Pesticide Residues Base on iScreen Printed Carbon Electrode/i(SPCE)-Bovine Serum Albumin (BSA)

         

摘要

The maximum level of organophosphate pesticide residues in rice is 0.1 mg/kg and 0.5 mg/kg in vegetables. The control of pesticide residues in agricultural products required a method of analysis quickly and accurately. The research developed a biosensor for the detection of organophosphate pesticide residues in agricultural products. The research studied immobilized organophosphate hydrolase (OPH) mass and characterization of biosensor. The solution conductivity measurement in the conductivity cell consists of a 1 × 5 mm2 pair of electrodes screen printed carbon electrode (SPCE). The instrument is a converted local conductometer. From the results of study concluded that the optimum performance of the biosensor was obtained from the 105 μg OPH, at pH 8.5 with a response time of 45 seconds. In that condition the sensitivity of biosensor is 28.04 μS/ppm and 0.18 ppm detection limit and the maximum concentration of pesticide which can be measured is 1 ppm. Biosensors have been applied to measure pesticide residues in some vegetable samples.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信