首页> 外文学位 >Reliability of the Thoratec HeartMate II flow measurements and alarms in the presence of reduced or non-existent flow.
【24h】

Reliability of the Thoratec HeartMate II flow measurements and alarms in the presence of reduced or non-existent flow.

机译:在流量减少或不存在的情况下,Thoratec HeartMate II流量测量和警报的可靠性。

获取原文
获取原文并翻译 | 示例

摘要

The most recognized risk associated with the HeartMate II is thrombosis. In the presence of developing clot, the HeartMate II Display Monitor and System Controller senses a decrease in pump flow and is accompanied by audible and visual alarms when flow rates drop below a certain threshold; however, when flow is completely inhibited, the Display Monitor and System Controller both fail to indicate that flow has reduced to zero and does not produce any corresponding alarms. To test the efficacy of the HeartMate II alarms, the Donavan Mock was used to simulate the hemodynamics of a typical heart failure patient. The hemodynamics were then improved by the addition of the HeartMate II LVAS. Partially occluding the inflow and outflow of the HeartMate II did display changes in flow and presented with alarms when appropriate; however, complete occlusions of the device failed to produce any alarms or accurate changes in flow.
机译:与HeartMate II相关的最公认的风险是血栓形成。当出现凝块时,HeartMate II显示监控器和系统控制器会感应到泵流量的减少,并且当流量降至某个阈值以下时,会伴随有声音和视觉警报。但是,当流量被完全禁止时,显示监视器和系统控制器都无法指示流量已减小到零,并且不会产生任何相应的警报。为了测试HeartMate II警报的功效,使用Donavan Mock模拟了典型心力衰竭患者的血液动力学。然后通过添加HeartMate II LVAS改善血液动力学。 HeartMate II的部分流入和流出确实显示了流量的变化,并在适当时显示了警报;但是,设备的完全堵塞无法产生任何警报或流量的准确变化。

著录项

  • 作者

    Hall, Seana G.;

  • 作者单位

    The University of Arizona.;

  • 授予单位 The University of Arizona.;
  • 学科 Health Sciences Medicine and Surgery.
  • 学位 M.S.
  • 年度 2013
  • 页码 27 p.
  • 总页数 27
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词