Electromagnetic interference (EMI) characteristics of implantable cardiac pacemakers and cardioverter defibrillators (ICDs) based on assessments of RFID interrogators are described in this paper. Examples of interrogators operating in various RF bands are tested to estimate EMI experienced by pacemakers and ICDs. The EMI test set-up and some experimental evaluation results for practical devices are shown. Observed characteristics of the EMI occurrence are discussed based upon the transmission radio wave specifications.
机译:本文介绍了基于RFID询问器评估的植入式心脏起搏器和心脏复律除颤器(ICD)的电磁干扰(EMI)特性。测试了在各种RF频段工作的询问器的示例,以估算起搏器和ICD所经历的EMI。显示了针对实际设备的EMI测试设置和一些实验评估结果。基于发射无线电波规范讨论了EMI发生的观察特性。