Plane wave fast color flow mode imaging: A parameter study

机译:平面波快速色流模式成像:参数研究

获取原文
获取原文并翻译 | 示例

摘要

A new Plane wave fast color flow imaging method (PWM) has been investigated, and performance evaluation of the PWM based on experimental measurements has been made. The results show that it is possible to obtain a CFM image using only 8 echo-pulse emissions for beam to flow angles between 45° and 75°. Compared to the conventional ultrasound imaging the frame rate is ~30-60 times higher. The bias, B_(est) of the velocity profile estimate, based on 8 pulse-echo emissions, is between 3.3% and 6.1% for beam to flow angles between 45° and 75°, and the standard deviation, σ_(est) of the velocity profile estimate is around 2% for beam to flow angles between 45° and 75° relative to the peak velocity, when the flow angle is known in advance. A study is performed to investigate how different parameters influence the blood velocity estimation. The results confirmed expectations for beam to flow angles between 45° and 75°. The parameter study shows that the PWM using Directional velocity estimation gives the best results using spatial sampling interval ≤ λ/10, correlation range ≥ 10λ, and number of directional signals ≥6. It is hereby shown that, by carefully choosing the set of parameters, PWM is feasible for fast CFM imaging with an acceptable bias and standard deviation.
机译:研究了一种新的平面波快速彩色流动成像方法(PWM),并基于实验测量结果对PWM的性能进行了评估。结果表明,对于束流角度在45°和75°之间的情况,仅使用8个回波脉冲发射就可以获得CFM图像。与传统的超声成像相比,帧速率高约30-60倍。速度分布估计的偏差B_(est)基于8个脉冲回波发射,射束与流角在45°至75°之间时的偏差为3.3%至6.1%,而标准偏差σ_(est)为当束流角相对于峰值速度在45°和75°之间时,如果事先知道流角,则速度分布估计约为2%。进行了一项研究,以研究不同的参数如何影响血流速度估计。结果证实了期望束流角在45°和75°之间。参数研究表明,使用方向速度估算的PWM在空间采样间隔≤λ/ 10,相关范围≥10λ和方向信号数≥6的情况下可获得最佳结果。由此表明,通过仔细选择参数集,PWM对于具有可接受的偏差和标准偏差的快速CFM成像是可行的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号