Evaluation of In Vitro Fatigue Properties of Biodegradable Mg-0.3at.Ca Alloy

机译:可生物降解的Mg-0.3at。%Ca合金的体外疲劳性能评估

获取原文

摘要

Recently, biodegradable bone fixation devices have been demanded when considering the patient's quality of life (QOL). During the fracture healing, the devices must support the repeated load due to daily performance. At the same time, surface of the magnesium devices was affected by body fluid. Thus in this research, in vitro fatigue properties of biodegradable Mg-0.3at.%Ca alloy was evaluated by using simulated body fluid. Though there was fatigue limit when the test was conducted under the ambient condition, it cannot be confirmed during the test in the simulated body fluid. Inspection of fracture surface revealed that crack propagated along the grain boundary after both the fatigue tests.
机译:近来,在考虑患者的生活质量(QOL)时,需要可生物降解的骨固定装置。在骨折愈合期间,由于日常性能,设备必须承受重复的载荷。同时,镁装置的表面受到体液的影响。因此,在这项研究中,通过使用模拟体液评估了可生物降解的Mg-0.3at。%Ca合金的体外疲劳性能。尽管在环境条件下进行测试时存在疲劳极限,但无法在模拟体液中进行测试。断裂表面的检查表明,在两次疲劳试验后,裂纹均沿晶界扩展。