首页> 外文会议>Optical Methods for Tumor Treatment and Detections: Mechanisms and Techniques in Photodynamic Therapy VII >New balloon catheter system used for PDT in the human urinary bladder: accuracy of light distribution
【24h】

New balloon catheter system used for PDT in the human urinary bladder: accuracy of light distribution

机译:用于人膀胱PDT的新型球囊导管系统:光分布的准确性

获取原文
获取原文并翻译 | 示例

摘要

Abstract: Photodynamic therapy (PDT) may provide a new approach for treatment of patients with superficial transitional carcinoma and carcinoma in situ of the bladder. The light applicator for the bladder wall (Rusch) is constructed as a balloon catheter with two concentric balloons. A new PDT applicator (Rusch) was assessed for the homogeneity and accuracy of irradiation during PDT. In an in-vitro experiment with 17 freshly harvested porcine bladders the fluence rate was measured locally with isotropic detectors. The results were compared to the light fluence detected by the PDT applicator. The increase of the fluence rate $beta inside the bladders due to back scattering ranged between 5.3 and 7.0 with an average of 6.2. Local variations of the fluence rate in the spherical bladders were also smaller than 15%. Therefore it is concluded, that a homogeneous and accurate irradiation during PDT is possible. Blood between the outer balloon and the bladder wall reduces the local fluence rate strongly and should to be avoided. Also larger air bubbles in the applicator can lead to an inhomogeneous light distribution. In regular application the presented new catheter system provides accurate and easy light dosimetry during PDT of the bladder. Attention had to be paid to a continuous flushing of the space between balloon and bladder wall in order to prevent the accumulation of urine and blood. To avoid a malfunction of the system and large errors in light dosimetry and application, it is advisable to monitor the measured light dosage and the shape of the balloon using ultrasonography during PDT. !6
机译:摘要:光动力疗法(PDT)可能为浅表性移行癌和膀胱原位癌的治疗提供一种新方法。用于膀胱壁的光施加器(Rusch)构造为具有两个同心球囊的球囊导管。对新的PDT施用器(Rusch)进行了PDT照射的均匀性和准确性评估。在一个用17个新鲜收获的猪膀胱进行的体外实验中,使用各向同性探测器对通量率进行了局部测量。将结果与PDT施加器检测到的光通量进行比较。由于向后散射,膀胱内的通量率$ beta的增加在5.3和7.0之间,平均为6.2。球形囊内注量率的局部变化也小于15%。因此可以得出结论,PDT期间均匀且准确的照射是可能的。外球囊和膀胱壁之间的血液会大大降低局部通量率,应避免使用。另外,涂抹器中的较大气泡也会导致光线分布不均匀。在常规应用中,提出的新型导管系统在膀胱PDT期间可提供准确而简便的光剂量测定。为了避免尿液和血液的积聚,必须注意持续冲洗气球和膀胱壁之间的空间。为避免系统出现故障以及光剂量测定和应用中的较大错误,建议在PDT期间使用超声检查来监测测得的光剂量和球囊的形状。 !6

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号