首页> 外文会议>Ophthalmic Technologies XV; Progress in Biomedical Optics and Imaging; vol.6 no.3 >Evaluation of porcine crystalline lenses in comparison with molded polymer gel lenses with an improved ex vivo accommodation simulator
【24h】

Evaluation of porcine crystalline lenses in comparison with molded polymer gel lenses with an improved ex vivo accommodation simulator

机译:通过改进的离体调节模拟器评估猪晶状体与模制聚合物凝胶状晶状体的比较

获取原文
获取原文并翻译 | 示例

摘要

The ex vivo accommodation simulator stretches crystalline lenses radially under controlled conditions, while monitoring the changes in zonular tension and lens diameter. A dual coaxial laser beam system was added for measuring changes in optical power as the lens is stretched. The EVAS instrument was fully characterized and calibrated. The results from 3 porcine eyes will be presented. Several crystalline lenses, similar in shape and size to porcine lenses, were cast molded in silicone gels of different stiffness. These lenses were evaluated using the EVAS instrument and custom made lens holding fixtures. With their well defined shape and material properties, results from these measurements will be useful for cross-validation of finite element models.
机译:体外调节模拟器在受控条件下径向拉伸晶状体,同时监测小带张力和晶状体直径的变化。添加了双同轴激光束系统,用于测量透镜拉伸后的光焦度变化。 EVAS仪器已经过全面表征和校准。将展示3只猪眼的结果。在不同硬度的硅凝胶中浇铸成型了数个晶体透镜,其形状和大小与猪晶状体相似。使用EVAS仪器和定制的镜片固定装置对这些镜片进行了评估。这些测量的结果具有明确定义的形状和材料特性,将对有限元模型的交叉验证很有用。