首页> 外文会议>International Conference on Biomedical Engineering and Informatics >Numerical Simulation of X-ray in-line Phase-contrast Imaging
【24h】

Numerical Simulation of X-ray in-line Phase-contrast Imaging

机译:X射线在线相位对比度成像的数值模拟

获取原文

摘要

A deterministic model based on Fresnel-Kirchhoff diffraction is proposed here to simulate X-ray in-line phase-contrast imaging. At first, a breast tissue-like sample with sharp edge is illuminated under our model by using a focal spot and polychromatic source and a finitely spatial resolution detector. The lateral coherence length and Gauss distribution function are introduced into our simulation model, in view of the finite source and detector. Then, experimental results were showed and compared with the simulation results. The great consistency between simulations and experiments shows validity of our simulation model in practical application.
机译:这里提出了一种基于FRESNEL-KIRCHOFF衍射的确定性模型,以模拟X射线在线相位对比度成像。首先,通过使用焦点和多色源和有限的空间分辨率检测器,在模型下照射具有尖锐边缘的乳房组织样样品。考虑到有限源和检测器,引入了横向相干长度和高斯分布函数。然后,显示实验结果并与模拟结果进行比较。模拟与实验之间的巨大一致性显示了我们在实际应用中的仿真模型的有效性。