We have investigated imaging characteristics of full-field OCT (FF-OCT) using a LED and a super-continuum (SC) laser with a wavelength-tunable filter. The wavelength selectivity in visible range enables to take spectroscopic tomographic images. It has been considered that the application of the LED-OCT system might be restricted due to the low spatial coherence and illuminating power. Comparative study of imaging characteristics of the OCT images taken the RGB-visible LED and the SC laser demonstrates superiority of the LED illumination in reconstruction of detail structures.
机译:我们使用带有波长可调滤波器的LED和超连续(SC)激光来研究全场OCT(FF-OCT)的成像特性。可见范围中的波长选择性使得能够采用光谱性断层图像。已经认为,由于空间相干性低和照明功率,LED-OCT系统的应用可能会受到限制。 OCT图像的成像特性的比较研究拍摄RGB可见LED和SC激光器在细节结构的重建中展示了LED照明的优越性。