【24h】

Evaluate the performance of spectroradiometer using BCRA tiles

机译:使用BCRA瓷砖评估光谱仪的性能

获取原文

摘要

Color stability is the key goal during the process of color dissemination and replication. The results of color consistent can be influenced by many factors. In this project, we focused on the error caused by measurement instrument, i.e. spectroradiometer. According to the radiance data collected from six British Ceramic Research Association tiles under two different standard illuminants, i.e. illuminant A and illuminant CWF, the tristimulus value of different colors under different conditions of illuminants were calculated and used for computing the mean color different from mean of the six tiles. Last, the performance of stability and consistency of the spectroradiometer was evaluated. The result showed that the spectroradiometer under standard illuminant CWF outperformed than that under standard illuminant A.
机译:颜色稳定性是颜色传播和复制过程中的关键目标。 颜色一致的结果可能受到许多因素的影响。 在该项目中,我们专注于测量仪器引起的误差,即光谱仪。 根据从六个英国陶瓷研究协会瓷砖收集的辐射数据,即在两种不同的标准光源,IE发光体A和发光体CWF,计算不同颜色在不同的照明条件下的三刺激度值,并用于计算与平均值不同的平均颜色 六块瓷砖。 最后,评估了光谱仪的稳定性和一致性的性能。 结果表明,标准光源CWF下的光谱辐射计比标准光源A在标准光子A下方。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号