首页> 外文会议>International Conference on Biomedical Engineering in Vietnam >Application of Transillumination Imaging to Injection Assist System
【24h】

Application of Transillumination Imaging to Injection Assist System

机译:过敏成像在注射辅助系统中的应用

获取原文

摘要

A technique of near-infrared (MR) trans-illumination imaging was applied to develop an injection assist system. We have developed a simultaneous observation system for a body surface image and a transillumination image of blood vessels. The system consists of a high-intensity and low-leak light source, near-infrared CMOS camera and a small and light one-eye head mounted display. In transillumination imaging, we could visualize the veins in the adult forearm with the thickness of as much as 67 mm. A model phantom of the veins in the forearm was fabricated with the blood-contained tubes buried in pork thigh meat. The blood tube was hardly observable with a naked eye. Using the developed system, we could identify the blood tube in the NIR transillumination image, and could extract the blood successfully with the assist of the NIR image.
机译:应用近红外(MR)跨辐照成像技术进行喷射辅助系统。我们已经开发了一种用于体表图像的同时观察系统和血管的过敏图像。该系统由高强度和低泄漏光源,近红外CMOS相机和小而轻的单眼镜安装显示器组成。在过敏成像中,我们可以使成年前臂中的静脉视为67mm的成人前臂。用埋在猪肉大腿肉中埋藏的血液含有管的血管中静脉模型幻影。用肉眼难以观察到血管。使用开发系统,我们可以识别NIR过敏图像中的血管,并可以通过NIR图像的辅助成功提取血液。