首页> 外文会议>Conference on Hard X-Ray and Gamma-Ray Detector Physics and Penetrating Radiation Systems >Feasibility Studies of an EMCCD-based Beta Imaging Probe for Radioguided Thyroid Surgery
【24h】

Feasibility Studies of an EMCCD-based Beta Imaging Probe for Radioguided Thyroid Surgery

机译:基于EMCCD的腺泡成像探针的可行性研究对辐射型甲状腺手术

获取原文

摘要

We are developing a probe for image-guided surgery of cancer to be used in conjunction with traditional beta emitting radiopharmaceuticals such as I~(131) and F~(18)-FDG. This device is based on a combination of two novel technologies, a microcolumnar film scintillator, CsI(Tl) and low-noise high sensitivity Electron-Multiplying CCD (EMCCD). The former allows high spatial resolution nuclear imaging and the latter facilitates detection of signal with significantly higher SNR than conventional CCDs by the virtue of internal signal gain built into its readout register. CsI(Tl) is bonded to the EMCCD via a flexible coherent fiberopic cable for easy handling. Due to its high sensitivity the probe is capable of functioning in real time providing high spatial resolution nuclear images for precise detection, delineation and excision of tumors. The evaluation of the probe using standard clinical phantoms as well as the operational data obtained on swine models and in clinical surgery will be presented.
机译:我们正在开发癌症的图像引导手术探头,与传统的β发射放射体放射性药物一起使用,例如I〜(131)和F〜(18)-FDG。该设备基于两种新颖技术,微校长薄膜闪烁体,CSI(TL)和低噪声高灵敏度电子 - 乘以CCD(EMCCD)的组合。前者允许高空间分辨率核成像,并且后者通过内部信号增益内置在其读出寄存器中的内部信号增益来促进与传统CCD显着更高的SNR的信号。 CSI(TL)通过柔性相干纤维电缆粘合到EMCCD,以便于处理。由于其高灵敏度,探头能够实时运作,提供高空间分辨率核图像以精确检测,描绘和肿瘤切除。将介绍使用标准临床模型的探针以及在猪模型和临床手术中获得的操作数据的评价。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号