首页> 外文会议>International Symposium on Heat Transfer Enhancement and Energy Conservation >DESIGN AND CALIBRATION OF A MINIATURE THIN-FILM HEAT-FLUX GAUGE
【24h】

DESIGN AND CALIBRATION OF A MINIATURE THIN-FILM HEAT-FLUX GAUGE

机译:微型薄膜热通量计的设计与校准

获取原文

摘要

A transient thin-film heat-flux gauge is designed and fabricated on silex substrate. It consists of a thermal resistance layer and a differential thermopile composed of twenty-eight pairs of Cu-Ni thermocouples, which measures the difference of temperature between the two sides of the thermal resistance layer. Calibration and testing of the gauge is carried out. During the steady-state test, the time constant of the heating period was 90s. The frequency response of the heat-flux gauge is measured in the frequency domain using a reflector-combined Halogen lamp and a chopper. It is found that the thin-film heat-flux gauge has a dynamic frequency response of 330Hz.
机译:在Silex基板上设计和制造了瞬态薄膜热通量计。它由热阻层和由二十八对Cu-Ni热电偶组成的差动热电堆组成,其测量热阻层的两侧之间的温度差。进行校准和测量标准的测试。在稳态测试期间,加热周期的时间常数为90秒。使用反射器组合的卤素灯和斩波器在频域中测量热通量计的频率响应。发现薄膜热通量计具有330Hz的动态频率响应。