This paper reports the micro-power telemetry unit that can sense pressure, humidity, and bio-potentials, for the study of micro-package techniques for implantable biomedical micro-systems used in research, diagnosis and therapy. The micro-package study is aimed to develop and demonstrate the micro-package technology for micro-implants to have an implant life of 0.2 to 2.0 years. The telemetry device consumes < 1.0 μA at 3 volts, with a volume < 0.2 cm~3 including lithium rechargeable battery, and has RF links to charge Li battery and to receive external on-off commands. The device output pulse width changes with a capacitive pressure sensor; and the pulse period changes with an external resistor as the humidity sensor or the bio-potential signals in series with the resistor. The first model of telemetry units, packaged with silicon and epoxy only, lasted more than 70 days in 40°C saline solution. The RF recharging and on-off control circuits was tested, and the lifetime of second generation telemetry unit are being evaluated.
机译:本文报道了可以感测压力,湿度和生物电位的微功率遥测单元,用于研究研究,诊断和治疗的可植入生物医学微型系统的微包技术。微包装研究旨在开发和证明微包装技术,用于微型植入物,使植入物寿命为0.2至2.0岁。遥测装置在3伏特下消耗<1.0μA,体积<0.2cm〜3,包括锂电电池,并具有充电Li电池的RF链路,并接收外部开关命令。设备输出脉冲宽度随电容式压力传感器而变化;并且脉冲周期随着湿度传感器的外部电阻或与电阻串联的生物电位信号而变化。遥测单元的第一个型号,仅用硅和环氧树脂包装,在40℃盐水溶液中持续超过70天。测试RF再充电和开关控制电路,正在评估第二代遥测单元的寿命。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号