首页> 外文会议>Conference on physics of medical imaging >Volume of Interest CT implemented with a dynamic bowtie filter
【24h】

Volume of Interest CT implemented with a dynamic bowtie filter

机译:使用动态领结滤波器实现的兴趣量CT

获取原文

摘要

Often in clinical CT imaging situations, a clinician only requires a small volume of interest (VOI) within a CT slice to be imaged. Unfortunately, due to some fundamental constraints in CT image reconstruction, the entire slice must be irradiated even when a small VOI is all that is clinically required. This produces excess dose for the patient, excess scatter which degrades image quality and increases clinician dose, and may effect patient care because clinicians may limit their use of CT in the interventional setting due to the previous two factors. This paper shows experimental results in which a digital beam attenuator (DBA) is used to modulate the patient dose as a function of view and gantry angle to enhance the SNR in a small VOI relative to the surrounding areas. It was found that the noise was lower by 50% inside the VOI for DBA enabled VOI. This promising result was obtained while providing an incident fluence reduction of 13 times on average for all regions not corresponding to the VOI relative to a non-VOI scan.
机译:通常,在临床CT成像情况下,临床医生仅需要在CT切片内进行少量成像(VOI)即可成像。不幸的是,由于CT图像重建中的一些基本限制,即使在临床上仅需要少量VOI的情况下,也必须对整个切片进行辐照。这给患者带来了过量的剂量,过量的散射降低了图像质量并增加了临床医生的剂量,并且可能会影响患者的护理,因为由于前两个因素,临床医生可能会限制他们在介入治疗中使用CT。本文显示了实验结果,其中使用数字束衰减器(DBA)来根据视角和机架角度调节患者剂量,以增强相对于周围区域的小VOI中的SNR。发现启用DBA的VOI内部的VOI噪声降低了50%。相对于非VOI扫描,在不对应VOI的所有区域平均使入射通量降低13倍的同时,获得了这一有希望的结果。