首页> 外文会议>World congress on medical physics and biomedical engineering >Evaluation of intra-aneurismal hemodynamics after stent placement in a pulsatile side wall model using laser Doppler anemometry
【24h】

Evaluation of intra-aneurismal hemodynamics after stent placement in a pulsatile side wall model using laser Doppler anemometry

机译:使用激光多普勒式气旋脉动侧壁模型在脉动侧壁模型支架置入后的运动内血流动力学评估

获取原文

摘要

Treatment of intracranial aneurysms by means of hemodynamic modification with the use of novel stent designs is burgeoning area of research. Despite promising results with recently approved implants there is no reliable method to predict long term success of this technique yet. We used Laser Doppler anemometry (LDA) to evaluate the influence of different stent designs on intra-aneurismal flow in a pulsatile aneurysm model. The effective reduction of flow was shown to be negligible with standard self expanding stents and depended on both the stent design and appropriate positioning of dedicated flow diverters. Our results show that Laser Doppler anemometry is a feasible method to quantify intra-aneurismal flow and flow reduction efficacy of stents in vitro.
机译:利用新型支架设计的血流动力学改性治疗颅内动脉瘤是蓬勃发展的研究领域。尽管较近期批准的植入物具有有前途的结果,但没有可靠的方法来预测这种技术的长期成功。我们使用了激光多普勒式气旋(LDA)来评估不同支架设计对脉动动脉瘤模型中的运动内流动的影响。有效减少流量与标准自膨胀支架易于计量,并依赖于支架设计和专用流动分流器的适当定位。我们的研究结果表明,激光多普勒风光测定法是一种可行的方法,用于量化体外支架的内胸内流量和流量减少效果。