In this paper, we propose new subdural electrode contacts to improve signal-to-noise ratio (SNR) for intracerebral electroencephalographic (icEEG) recording. We describe the proposed electrodes material composition, fabrication process, electro-chemical impedance spectroscopy analyses, and in vitro comparative studies with commercial electrodes (micro-electrodes-MRE, depth electrodes -DPE, and subdural electrodes -SDE). The proposed subdural electrode contacts are made of Platinum (Pt) or Gold (Au). Compared to the DPE/SDE and MRE, the proposed Pt or Au electrodes feature reduced impedances of ∼116 kΩ and ∼8.89 MΩ respectively. In vivo icEEG recordings for three weeks in an adult Sprague Dawley rat demonstrated signal stability, 50% noise reduction and up to 3 dB SNR improvement. Following the long-term icEEG recording, brain histological result showed no abnormal tissue reaction in underlying cortex.
机译:在本文中,我们提出了一种新的硬膜下电极触点,以改善脑电脑电图(icEEG)记录的信噪比(SNR)。我们描述了拟议的电极材料组成,制造工艺,电化学阻抗谱分析以及与商用电极(微电极-MRE,深度电极-DPE和硬膜下电极-SDE)的体外比较研究。建议的硬膜下电极触点由铂(Pt)或金(Au)制成。与DPE / SDE和MRE相比,建议的Pt或Au电极的阻抗分别降低了约116kΩ和约8.89MΩ。一只成年Sprague Dawley大鼠的体内icEEG录音记录了三周,显示出信号稳定性,降低了50%的噪声以及提高了3 dB的SNR。经过长期的icEEG记录,大脑组织学结果显示基础皮层无异常组织反应。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号