Maximal ratio combining for respiratory effort extraction from pressure sensor arrays

机译:最大比率组合,可从压力传感器阵列中提取呼吸力

获取原文

摘要

Respiratory effort signals may be measured using pressure sensors placed below a mattress. When few of the sensors include the respiratory effort signal or the signal power is low, output respiratory effort signal quality is low. This paper examines current signal combining methods used in pressure sensor array breathing estimation and proposes a maximal ratio combining method to produce the best estimate in these conditions. Data was collected from two participants and the effects of low signal power and low signal availability were simulated by corrupting the recordings. The results show that the proposed method offers 2.5 dB higher signal to noise ratio in the presence of substantial noise and good correlation (> 0.92) to gold standard respiration bands.
机译:可以使用放置在床垫下方的压力传感器来测量呼吸力信号。当很少有传感器包含呼吸力信号或信号功率低时,输出的呼吸力信号质量会降低。本文研究了用于压力传感器阵列呼吸估计的电流信号合并方法,并提出了一种最大比率合并方法以在这些条件下产生最佳估计值。从两名参与者收集数据,并通过破坏录音来模拟低信号功率和低信号可用性的影响。结果表明,在存在大量噪声且与金标准呼吸带具有良好相关性(> 0.92)的情况下,该方法可提供高出2.5 dB的信噪比。