A technique for calibrating the spectral radiance responsivity of the CCD imaging spectrometer with an internally illuminated integrating sphere is described. The spectral radiance of the integrating sphere is obtained by two steps. Firstly, a Spectralon panel diffuser and an ultraviolet spectrometer are combined into a new spectral radiometer which transfers the spectral irradiance of a NIST standard of spectral irradiance to that of the receiving aperture of the integrating sphere. Subsequently, the spectral radiance of the integrating sphere is derived from heat transfer theory for Lambertian radiators. The overall uncertainty of determining the spectral radiance of the integrating sphere is ±2.3%. On the basis of known spectral radiance, the radiance calibration of an available Czerny-Turner imaging spectrometer in our laboratory has been completed in 200-400nm with an uncertainty of about ±2.7%.
机译:描述了一种用于校准带有内部照明的积分球的CCD成像光谱仪的光谱辐射响应度的技术。积分球的光谱辐射度通过两个步骤获得。首先,将Spectralon平板扩散器和紫外线光谱仪组合成一个新的光谱辐射仪,该光谱仪将NIST标准的光谱辐照度的光谱辐照度转移到积分球的接收孔径。随后,根据朗伯辐射器的传热理论,得出了积分球的光谱辐射率。确定积分球光谱辐射度的总不确定度为±2.3%。根据已知的光谱辐射度,我们实验室中可用的Czerny-Turner成像光谱仪的辐射度校准已在200-400nm范围内完成,不确定度约为±2.7%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号