【24h】

Extending the range for precision AM noise measurements

机译:扩展了用于精确AM噪声测量的范围

获取原文

摘要

Developed as an extension of the existing NIST phase noise measurement system, the AM measurement system presently covers carrier frequencies from 5 MHz to 70 GHz. Our new technique relies on a wideband AM modulator to calibrate frequency dependent errors. The system, utilizing a double-balanced mixer as the detector, works in two passes. In the first pass, the modulator is used to sweep a known modulation level across the Fourier frequencies of interest. This level is detected by three different spectrum analyzers that cover Fourier frequencies from 1 Hz to 1 GHz. The data is then compiled into a calibration curve that is used to correct frequency dependent errors in the noise data measured in the second pass. The entire system is computer controlled via the GPIB bus enabling measurements to be made with minimal user intervention.
机译:作为现有NIST相位噪声测量系统的扩展,AM测量系统目前覆盖5 MHz至70 GHz的载波频率。我们的新技术依靠宽带AM调制器来校准频率相关的误差。该系统利用双平衡混频器作为检测器,可分两遍工作。在第一遍中,使用调制器在感兴趣的傅立叶频率上扫描已知的调制电平。可以通过覆盖1 Hz至1 GHz傅立叶频率的三个不同的频谱分析仪来检测此电平。然后将数据编译成校准曲线,该校准曲线用于校正第二遍测量的噪声数据中与频率有关的误差。整个系统通过GPIB总线进行计算机控制,从而可以在最少的用户干预下进行测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信