A clinical study of short-term sternal photoplethysmography: Recordings from patients with obstructive airways diseases

机译:短期胸骨光体积描记术的临床研究:阻塞性气道疾病患者的记录

获取原文

摘要

Traditionally, measurements of the oxygen saturation (SpO2) has been confound to the extremities. In this study, we therefore investigated the possibility for reliable estimation of clinically relevant SpO2 levels from photoplethysmography (PPG) obtained on the sternum of patients with obstructive airway diseases. We initiated the study with a calibration of a prototype sternal PPG sensor. In accordance with the ISO 80601-2-61:2011 guidelines, the calibration was conducted as a controlled desaturation study. We obtained a calibration accuracy of 1.75% which is well within the clinically and commercially accepted range. We then compared the SpO2 levels simultaneously obtained from the sternal PPGs and a commercially available finger pulse oximeter on 28 admitted patients with either asthma or Chronic Obstructive Pulmonary Disease (COPD). The Pearson correlation between the SpO2 levels estimated from the two body locations was found to be 0.89 (p<;0.05) and the mean system bias was only 0.052% with upper and lower limits of agreement of 2.5% and -2.4%, respectively. This finding is very promising for the future design of new sternum based patch technologies that might be able to provide continuous estimates of the SpO2 levels on critically or chronically ill patients.
机译:传统上,四肢的氧饱和度(SpO2)的测量结果很混乱。因此,在这项研究中,我们调查了从阻塞性气道疾病患者的胸骨上获得的光体积描记法(PPG)可靠估计临床相关SpO2水平的可能性。我们通过对原型胸骨PPG传感器进行校准来启动研究。根据ISO 80601-2-61:2011指南,将校准作为受控的去饱和研究进行。我们获得的校准精度为1.75%,完全在临床和商业上可接受的范围内。然后,我们比较了从胸骨PPG和市售的手指脉搏血氧仪同时获得的SpO2水平,对28名哮喘或慢性阻塞性肺疾病(COPD)的入院患者进行了比较。从两个身体部位估计的SpO2水平之间的Pearson相关性被发现为0.89(p <; 0.05),平均系统偏差仅为0.052%,一致性的上限和下限分别为2.5%和-2.4%。这一发现对于未来基于新的基于胸骨的贴片技术的设计很有希望,该技术可能能够对重症或慢性病患者的SpO2水平进行连续估计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号