首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Simulation of Transcatheter Aortic Valve Replacement in patient-specific aortic roots: Effect of crimping and positioning on device performance
【24h】

Simulation of Transcatheter Aortic Valve Replacement in patient-specific aortic roots: Effect of crimping and positioning on device performance

机译:特定患者主动脉根中经导管主动脉瓣置换的模拟:压接和定位对装置性能的影响

获取原文

摘要

Calcific aortic valve disease (CAVD) is a cardiovascular condition that causes the progressive narrowing of the aortic valve (AV) opening, due to the growth of bone-like deposits all over the aortic root (AR). Transcatheter aortic valve replacement (TAVR), a minimally invasive procedure, has recently become the only lifesaving solution for patients that cannot tolerate the standard surgical valve replacement. However, adverse effects, such as AR injury or paravalvular leakage (PVL), may occur as a consequence of a sub-optimal procedure, due to the presence of calcifications in situ. Additionally, the crimping required for delivering the valve via stenting may damage the valve. The aim of the present study is to comparatively assess the crimping mechanics of the commercialized Edwards SAPIEN valve and an alternative polymeric valve (Polynova, Inc) and to evaluate the effect of different TAVR deployment positions using patient-specific numerical models. The optimal deployment location for achieving better patient outcomes was calculated and based on the interactions between the TAVR stent and the native AR. Results demonstrated that the Polynova valve withstands the crimping process better than the SAPIEN valve. Furthermore, deployment simulations showed the role that calcifications deposits may play in the TAVR sub-optimal valve anchoring to the AV wall, leading to the presence of gaps that result in PVL.
机译:钙化性主动脉瓣疾病(CAVD)是一种心血管疾病,由于整个主动脉根部(AR)均出现骨样沉积物,导致主动脉瓣(AV)开口逐渐变窄。经导管主动脉瓣置换术(TAVR)是一种微创手术,最近已成为无法忍受标准外科瓣膜置换术的患者的唯一救生解决方案。然而,由于原位钙化的存在,次优手术可能会导致不良反应,例如AR损伤或瓣周漏(PVL)。另外,通过支架输送阀所需的压接可能损坏阀。本研究的目的是比较评估商业化的Edwards SAPIEN瓣膜和替代聚合物瓣膜(Polynova,Inc)的压接机理,并使用特定于患者的数值模型评估不同TAVR部署位置的影响。根据TAVR支架与天然AR之间的相互作用,计算出用于实现更好患者预后的最佳部署位置。结果表明,Polynova阀门比SAPIEN阀门更能承受压接过程。此外,部署模拟表明钙化沉积物可能在锚定于AV壁的TAVR次优瓣膜中发挥作用,导致存在导致PVL的间隙。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号