Novel All-State Heavy Metal Sensors and Analysis Instrument

机译:新型全态重金属传感器和分析仪器

获取原文

摘要

Heavy metals, which accumulate into human body and cause many diseases for never degradation, can lead to undesirable hazard to human being. Novel all-state mercury ion selective electrode (Hg-ISE) and Cadmium ion selective electrode (Cd-ISE) are prepared and analysis instrument based on the two electrodes is developed. The detection limit of Hg-ISE is 3.31×10{sup}(-6) mol/L with the scope of 29.3 mV/decade in the linearity of calibration curve. The detection limit of Cd-ISE is 5.56×10{sup}(-6) mol/L with the scope of 27.5 mV/decade. Two electrodes with simple structure are convenient to be used. The water sample to measure is only needed to 10 ml. When the concentration of heavy metal is higher, it needs only several seconds that response voltage is able to keep stable. Test results show the two electrodes have advantages of high chemical stabilization and selectivity, repetitiveness of voltage, corrosion resistance. The analysis instrument has features of short detection time, sufficient function, convenient interface and stable performance.
机译:重金属,这些积聚成人体和引起许多疾病为从不退化,可导致不希望的危害人类。新颖的全状态汞离子选择性电极(汞ISE)和镉离子选择性电极(CD-ISE)制备以及基于所述两个电极分析仪被显影。汞-ISE的检测限是3.31×10 {SUP}( - 6)摩尔/ L的29.3毫伏/十年在校准曲线的线性范围。镉-ISE的检测限是5.56×10 {SUP}( - 6)摩尔/ L为27.5毫伏/十倍的范围。两个电极结构简单方便使用。水样品以测量时才需要至10ml。当重金属的浓度较高时,它只需要几秒钟即响应电压是能够保持稳定。测试结果表明,该两个电极具有高的化学稳定性和选择性,电压的重复性,耐腐蚀的优点。所述分析仪器具有检测时间短,充分的功能,接口方便,性能稳定的特性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号