首页> 外文会议>World multi-conference on systemics, cybernetics and informatics >An Implantable Cardiovascular Pressure Monitoring System with On-chip Antenna and RF Energy Harvesting
【24h】

An Implantable Cardiovascular Pressure Monitoring System with On-chip Antenna and RF Energy Harvesting

机译:具有片上天线和RF能量收集的植入心血管压力监测系统

获取原文

摘要

An implantable wireless system with on-chip antenna for cardiovascular pressure monitor is studied. The implantable device is operated in a batteryless manner, powered by an external radio frequency (RF) power source. The received RF power level can be sensed and wirelessly transmitted along with blood pressure signal for feedback control of the external RF power. The integrated electronic system, consisting of a capacitance-to-voltage converter, an adaptive RF powering system, an RF transmitter and digital control circuitry, is simulated using a TSMC 0.18 μm CMOS technology. The implanted RF transmitter circuit is combined with a low power voltage-controlled oscillator resonating at 5.8 GHz and a power amplifier. For the design, the simulation model is setup using ADS and HFSS software. The dimension of the antenna is 1× 0.6 × 4.8 mm~3 with a 1 × 0.6 mm~2 on-chip circuit which is small enough to place in human carotid artery.
机译:研究了具有用于心血管压力监测器的片上天线的可植入无线系统。可植入设备以无电射方式操作,由外部射频(RF)电源供电。接收的RF功率电平随着用于反馈控制的外部RF功率的反馈控制,可以感测和无线传输。使用TSMC0.18μmCMOS技术模拟由电容到电压转换器,自适应RF供电系统,RF发射器和数字控制电路组成的集成电子系统。植入的RF发射器电路与低功率电压控制振荡器组合在5.8GHz和功率放大器中谐振。对于设计,使用ADS和HFSS软件设置仿真模型。天线的尺寸为1×0.6×4.8mm〜3,其中1×0.6mm〜2片上电路,足够小,放置在人颈动脉中。