Microfluidic device to perform impedometric detection of Activated Partial Thromboplastin Time of blood

机译:微流控设备用于血液中活化部分凝血活酶时间的阻抗测定

获取原文

摘要

The Activated Partial Thromboplastin Time (APTT) test is performed on a large scale, in laboratories to monitor the functioning of intrinsic and common pathways of blood coagulation cascade and the concentration of anticoagulants such as Heparin among patients. A PDMS based microfluidic device has been fabricated and tested to perform APTT as a point-of-care device using whole blood samples by detecting the change in electrical impedance of blood during coagulation. The two chambered PDMS device integrated with microelectrodes is pre-coated with APTT reagent and heparin respectively for better repeatability. The results indicate a repeatable pattern of change in whole blood impedance during coagulation and confirm the theoretical explanation by Hartert [1]. The devices were also tested for their sensitivity to different concentrations of heparin and found to exhibit a favorable increase in coagulation time for increasing heparin concentration.
机译:活化部分凝血活酶时间(APTT)测试在实验室中进行,以监测患者体内凝血级联的内在和常见途径的功能以及抗凝剂(例如肝素)的浓度。已经制造并测试了基于PDMS的微流体设备,可以通过检测凝血过程中血液的电阻抗变化来使用全血样本作为AP-Point-Point现场护理设备执行APTT测试。集成有微电极的两室PDMS装置分别预先涂有APTT试剂和肝素,以提高重复性。结果表明凝血过程中全血阻抗变化的可重复模式,并证实了Hartert [1]的理论解释。还测试了这些装置对不同浓度肝素的敏感性,发现对于增加肝素浓度,凝血时间表现出有利的增加。