首页> 外文会议>2011 IEEE 24th International Conference on Micro Electro Mechanical Systems >Fabrication of bioprobe integrated with hollow nanoneedle for cellular function analysis
【24h】

Fabrication of bioprobe integrated with hollow nanoneedle for cellular function analysis

机译:集成空心纳米针的生物探针的细胞功能分析

获取原文

摘要

In order to realize cellular function analysis in a single living cell, we have been developing a newly designed probe for atomic force microscope (AFM), named “bioprobe”, which is integrated with a hollow silicon dioxide (SiO2) nanoneedle. The probe will be capable not only of performing AFM measurements but also of introducing desired biomolecules (DNA, proteins, etc.) into living cells and extracting biomolecules expressed in the cells. In this paper, we demonstrated the ability of the fabricated bioprobe to capture AFM images of living HeLa cells under near-physiological conditions. Moreover, the sharp-tipped SiO2 nanoneedle can be inserted into the cells without significant cell damage. In order to examine the possibility of liquid delivery, water ejection tests were also carried out. As a result, an extremely small amount of water can be ejected through a nanoneedle.
机译:为了在单个活细胞中实现细胞功能分析,我们正在开发一种新设计的原子力显微镜(AFM)探针,名为“ bioprobe”,该探针与中空二氧化硅(SiO 2 )nanoneedle。该探针将不仅能够执行AFM测量,而且还能将所需的生物分子(DNA,蛋白质等)引入活细胞并提取在细胞中表达的生物分子。在本文中,我们证明了人造生物探针在接近生理条件下捕获活HeLa细胞的AFM图像的能力。此外,可以将尖锐的SiO 2 纳米针插入细胞,而不会造成明显的细胞损伤。为了检查液体输送的可能性,还进行了喷水测试。结果,极少量的水可以通过纳米针喷出。