The development of a cylindrical-core fiber optic oxygen sensor for fast measurement of oxygen partial pressure (pO_2) is described. The fibre sensing element is based on a cylindrical-core waveguide structure formed by a polymer sensing film that contains immobilized Pt (II) complexes. The performance of the fibre optic oxygen sensor was evaluated using luminescence intensity measurement. To determine accurately the response time of the oxygen sensor, a test chamber was used to provide rapid changes in the partial pressure of oxygen. The result showed that the time response of this cylindrical-core fibre based optic oxygen sensor can be less than 0.2 second. The details of the sensor chemistry, experimental system and the results are reported.
机译:描述了一种用于快速测量氧气分压(pO_2)的圆柱芯光纤氧气传感器的开发。光纤传感元件基于由包含固定的Pt(II)配合物的聚合物传感膜形成的圆柱芯波导结构。使用发光强度测量来评估光纤氧气传感器的性能。为了准确确定氧气传感器的响应时间,使用了一个测试室来提供氧气分压的快速变化。结果表明,这种基于圆柱芯光纤的光学氧气传感器的时间响应可以小于0.2秒。报告了传感器化学,实验系统和结果的详细信息。