Imaging QA Program for Image Guided RadioTherapy (IGRT)

机译:影像引导放射治疗(IGRT)的影像质量检查程序

获取原文

摘要

State of the art radiation therapy devices offer the possibility to acquire images directly before the treatment of the patients. These images can be used to correct for the position of the patient or to recalculate dose on a changed patient anatomy.In this work a QA program for such an imaging modality is presented on the basis of a TomoTherapy Hi-Art II system. The QA program consists of weekly, monthly and biannual checks of the imaging device.The results of the performed QA measurements were all within the tolerances of the system.The imaging capability of the TomoTherapy Hi-Art II system is an excellent tool for image guided positioning of the patient. Furthermore it is possible to perform a reliable dose calculation on the basis of the acquired images.
机译:先进的放射治疗设备提供了在患者治疗之前直接获取图像的可能性。这些图像可用于校正患者的位置或重新计算已改变的患者解剖结构上的剂量。 在这项工作中,基于TomoTherapy Hi-Art II系统提出了一种用于这种成像方式的QA程序。质量检查程序包括对成像设备进行每周,每月和每年两次的检查。 进行质量检查的结果均在系统的公差范围内。 TomoTherapy Hi-Art II系统的成像能力是用于图像引导患者定位的出色工具。此外,可以基于所获取的图像执行可靠的剂量计算。