首页> 外文会议>World congress on medical physics and biomedical engineering;International congress of the IUPESM >Relative peripheral dose at IMRT- A comparison between Siemens Oncor and Artiste
【24h】

Relative peripheral dose at IMRT- A comparison between Siemens Oncor and Artiste

机译:IMRT的相对外周剂量-Siemens Oncor和Artiste的比较

获取原文

摘要

The aim of this work was to investigate how the relative peripheral dose (RPD) outside the primary treatment field changes in IMRT in connection with the use of either the 82-leaf MLC or the 160-leaf MLC from Siemens. 9-beam plans based on the CT data set of an acrylic phantom for 6MV photon energy were calculated for the Oncor and Artiste linacs in such a way as to get nearly identical DVHs for the artificial PTV- and OARs. Horizontal and vertical dose profiles inside the phantom were measured using thermoluminescent detectors TLD700 (rods). The measured dose values were normalized to lOGy median dose in the PTV. With increasing distance from the isocentre the RPD follows a slightly exponential decrease. On average, values for the 160MLC are 15% lower than those of the 82MLC up to a distance of 23.5cm from isocentre. This is attributed to the 2cm thicker MLC material which absorbes headscatter. Based on these observations it can be stated that the Artiste with the new MLC produces less peripheral dose and offers an improved treatment of patients with IMRT.
机译:这项工作的目的是研究与使用西门子的82叶MLC或160叶MLC一起使用时,IMRT中主要治疗范围以外的相对外周剂量(RPD)如何变化。为Oncor和Artiste直线加速器计算了基于丙烯酸模体的6MV光子能量的CT数据集的9束计划,从而获得了与人造PTV和OAR几乎相同的DVH。使用热发光检测器TLD700(棒)测量体模内部的水平和垂直剂量分布。将测量的剂量值标准化为PTV中的10Gy中值剂量。随着距等中心点距离的增加,RPD呈指数级下降。平均而言,在距等心点23.5厘米处,160MLC的值比82MLC的值低15%。这归因于2cm厚的MLC材料吸收了头部散射。基于这些观察结果,可以说使用新型MLC的Artiste产生的外周剂量更少,并为IMRT患者提供了更好的治疗方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号