Measurement of peroxyacetyl nitrate NO2 and NOx by using a gas chromatograph with a luminol-based detector

机译:使用带鲁米诺检测器的气相色谱仪测量过氧乙酰硝酸盐NO2和NOx

获取原文
获取原文并翻译 | 示例

摘要

Abstract: A commercial instrument for measuring PAN (peroxyacetyl nitrate), NO$- 2$/, and optionally, NOx, is described. It is based on a gas chromatograph (GC), and uses a Luminox$+R$/ luminol detector following conversion of the PAN to NO$-2$/. A detection limit (based on a peak height of 3 $MUL noise) of 30 pptV is achieved for injections made every five minutes. The instrument is computer driven and contains its own peak integrator and graphics output. No interferences have been identified, including chlorinated hydrocarbons, NO$-2$/ and ozone. However, the PAN analog peroxypropanyl nitrate (PPN) is not fully resolved from PAN. The maintenance interval is monthly. The calibration of the instruments is relatively easy: known mixing ratios of NO$-2$/ in air are used to calibrate the sample injection system and Luminox$+R$/ detector. The passage of PAN through the column and the subsequent conversion of PAN to NO$-2$/ has been found to be stable over times of one year, and is reproducible between instruments. Atmospheric measurements for PAN have been carried out for both very polluted environments and very clean environments. The results from a month long formal intercomparison between the Luminox$+R$/ PAN and two GCs equipped with electron capture detectors is presented. Those results demonstrate that the luminol based PAN detector can yield accurate measurements of ambient PAN concentrations. A method for calibrating PAN instruments in the field is described.!
机译:摘要:描述了一种商用仪器,用于测量PAN(过氧乙酰硝酸盐),NO $ -2 $ /和可选的NOx。它基于气相色谱仪(GC),在将PAN转换为NO $ -2 $ /之后使用Luminox $ + R $ /鲁米诺检测器。每五分钟进样可达到30 pptV的检出限(基于3 $ MUL噪声的峰高)。该仪器由计算机驱动,并包含其自己的峰值积分器和图形输出。没有发现干扰,包括氯代烃,-2美元和臭氧。但是,PAN模拟硝酸过氧丙酯(PPN)不能完全从PAN中分离出来。维护间隔为每月一次。仪器的校准相对容易:已知的空气中NO $ -2 $ /的混合比用于校准样品注射系统和Luminox $ + R $ /检测器。已发现PAN通过色谱柱的传递以及随后PAN转化为NO $ -2 $ /的稳定性在一年的时间内是稳定的,并且在仪器之间是可重现的。 PAN的大气测量已经在非常污染的环境和非常清洁的环境中进行了。给出了Luminox $ + R $ / PAN与两个配备电子捕获检测器的GC之间长达一个月的正式比对的结果。这些结果表明,基于鲁米诺的PAN检测器可以准确测量环境PAN浓度。描述了一种在现场校准PAN仪器的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号