Investigation on gPhone gravimeter-119 for gravity variations observation during the 10th International Comparison of Absolute Gravimeters (ICAG-2017)

机译:第10届国际重力仪国际比较(ICAG-2017)中用于观测重力变化的gPhone gravimeter-119的研究

获取原文
获取原文并翻译 | 示例

摘要

The 10th International Comparison of Absolute Gravimeters (ICAG-2017) was held in Changping campus of NationalInstitute of Metrology (NIM), China in October 2017. The observation of gravity variations using relative gravimetersplays an important role in absolute gravimeter comparison and the link of gravity reference value after comparison. Wecarried out a continuous observation of gPhone gravimeter-119 simultaneously alongside a superconductive gravimeteriGrav-012 for several months. The calibration factor of gPhone-119 is determined to be 0.99355±0.00004 with aprecision of 0.004%. When the observation time exceeds 33000 minutes, the calibration values and uncertainties tend tobe stable and the precision is better than 0.01%. The non-tidal gravity changes during ICAG-2017 recorded bygPhone-119 are analyzed. The tendency of gravity variations is roughly consistent with that recorded by iGrav-012. Theresult indicates that the peak-to-peak value of gravity changes is less than 1.5 μGal during the period of ICAG-2017.
机译:第十届国际绝对重力比较仪(ICAG-2017)于2017年10月在中国国家计量学会(NIM)昌平校区举行。使用相对重力仪观测重力变化在绝对重力仪的比较和重力参考值比较后的链接。我们同时使用超导重力仪\ r \ niGrav-012同时连续观察gPhone重力仪119数月。 gPhone-119的校准系数确定为0.99355±0.00004,精度为0.004%。当观察时间超过33000分钟时,校准值和不确定度趋于稳定,精度优于0.01%。分析了\ r \ ngPhone-119记录的ICAG-2017期间的非潮汐重力变化。重力变化的趋势与iGrav-012记录的趋势大致一致。结果表明在ICAG-2017期间重力变化的峰峰值小于1.5μGal。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号