The study presents the results of tests of electrochemical corrosion resistance of wires made of stainless steel X10CrNil8-8 for guides used in invasive cardiology. Measurements were made in artificial blood plasma solution by means of electrochemical tests system VoltaLab®PGP201. On the ground of the registered anodic polarisation curves, factors that characterise electrochemical corrosion resistance, i.e..: breakdown potential, polarisation resistance and corrosion current density, have been described. In order to evaluate phenomena that take place on the surface of the tested steel. Electrochemical Impedance Spectroscopy was employed. Measurements were made with the measurement system AutoLab PGSTAT 302N equipped with FRA2 (Frequency Response Analyser) module. During the tests, impedance spectra of the system were made and obtained data was matched to the equivalent circuit. Moreover, the effect of strain in the drawing process and the way of wire surface preparation on the resistance to electrochemical corrosion.
机译:该研究提出了X10CrNil8-8不锈钢制成的导线的抗电化学腐蚀性能的测试结果,这些导线可用于有创心脏病学。借助于电化学测试系统PGP201在人造血浆溶液中进行测量。基于所记录的阳极极化曲线,已经描述了表征电化学耐腐蚀性的因素,即:击穿电位,极化电阻和腐蚀电流密度。为了评估在被测钢表面上发生的现象。使用电化学阻抗谱。使用配备FRA2(频率响应分析器)模块的测量系统AutoLab PGSTAT 302N进行测量。在测试过程中,制作了系统的阻抗谱,并将获得的数据与等效电路匹配。此外,拉拔过程中的应变和导线表面制备方式对耐电化学腐蚀的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号