首页> 外文会议>Electrical measurements and instrumentation >INFLUENCE OF DVM'S INPUT PARAMETERS ON DVM-BASED HIGH-OHM RESISTANCE COMPARISON
【24h】

INFLUENCE OF DVM'S INPUT PARAMETERS ON DVM-BASED HIGH-OHM RESISTANCE COMPARISON

机译:DVM输入参数对基于DVM的高OHM电阻比较的影响

获取原文
获取原文并翻译 | 示例

摘要

Establishment and maintenance of resistance rnstandard traceability is one of the main tasks of Primary rnElectromagnetic Laboratory (PEL) within the Faculty of rnElectrical Engineering and Computing of the University of rnZagreb. For the purpose of accurate resistance standards rncomparison a digital voltmeter (DVM) based method has rnbeen designed and realized. When measuring resistance rnstandards greater than 10 MΩ capabilities of the method are rnrestricted due to influence of DVM's input parameters: input rnDVM's resistance, input offset current and input rncapacitance. Determination of DVM's input parameters and rntheir contribution to the overall high-ohm resistance rncomparison uncertainty has been presented in this paper. rnThe possibility of use of DVM-based method for gigaohm rnvalue resistors comparison has also been investigated.
机译:建立和维护电阻标准可追溯性是rnZagreb大学rn电磁工程与计算学院内主要rn电磁实验室(PEL)的主要任务之一。为了精确的电阻标准比较,已经设计并实现了一种基于数字电压表(DVM)的方法。当测量大于10MΩ的电阻标准时,由于DVM输入参数的影响,该方法的功能会受到限制:输入DVM的电阻,输入失调电流和输入电容。本文介绍了DVM输入参数的确定及其对整个高欧姆电阻不确定性的贡献。还研究了将基于DVM的方法用于千兆欧rnvalue电阻比较的可能性。