摘要

A system for ophthalmic surgery support has been developed in order to minimize the residual astigmatism due to the induced irregular shape of the cornea by corneal suture. The system projects 36 light spots, from LEDs, displayed in a precise circle at the lachrymal film of the examined cornea. The displacement, the size and deformation of the reflected image of these light spots are analyzed providing the keratometry and the circularity of the suture. Measurements in the range of 32D - 55D (up to 23D of astigmatism are possible to be obtained) and a self-calibration system has been designed in order to keep the system calibrated. Steel precision spheres have been submitted to the system and the results show 99% of correlation with the fabricant's nominal values. The system has been tested in 13 persons in order to evaluate its clinical applicability and has been compared to a commercial keratometer Topcon OM-4. The correlation factors are 0,92 for the astigmatism and 0.99 for the associated axis. The system indicates that the surgeon should achieve circularity ≥98% in order to do not induce astigmatisms over 3D.
机译:已经开发了用于眼科手术支持的系统,以最小化由于角膜缝线引起的角膜的不规则形状而导致的残留散光。该系统从LED投射36个光斑,精确地显示在被检角膜的泪膜上。分析这些光斑的反射图像的位移,大小和变形,从而提供角膜测量法和缝合线的圆度。测量范围为32D-55D(可能获得多达23D的像散),并且设计了自校准系统以保持系统校准。钢制精密球已提交给系统,结果显示与制造商的标称值具有99%的相关性。该系统已经在13个人中进行了测试,以评估其临床适用性,并已与商业角膜曲率计Topcon OM-4进行了比较。相关系数散光为0.92,关联轴为0.99。系统表明,外科医生应达到≥98%的圆度,以免在3D上引起散光。