Parametric Study of a Conical Applicator for 4 MHz HIFU Transducer for its Possible Application in Bone Tumour Ablation Therapy

机译:用于4 MHz HIFU换能器的锥形敷料器的参数研究可能在骨肿瘤消融治疗中的应用

获取原文

摘要

In thermal ablation therapy, high intensity focused ultrasound is used to concentrate the acoustic beams in a specific region. This ultrasonic energy concentration causes a rapid temperature increase in a short time. In some clinical applications, a truncated cone coupled to the transducer contains a propagation medium to allow the delivery of the therapy; besides, its shape points in the target direction. This paper presents a parametric study of a conical applicator for a 4 MHz HIFU transducer with 100 mm focal length for the evaluation of its use in bone tumors. The conical applicator was varied in radius and thickness from 4 -7.5 mm and 1 to 2 mm, respectively, in order to achieve similar acoustic distribution to the HIFU transducer without cone. Three different materials compatible with 3D printing were studied: polylactic acid, acrylonitrile butadiene styrene and low density polyethylene. Simulations were realized in COMSOL Multiphysics
机译:在热消融治疗中,高强度聚焦超声用于将声束集中在特定区域中。该超声能量集中导致在短时间内温度快速升高。在某些临床应用中,连接到换能器的截锥包含传播介质以允许治疗的传递。此外,其形状指向目标方向。本文介绍了一个焦距为100 mm的4 MHz HIFU换能器的锥形敷贴器的参数研究,以评估其在骨肿瘤中的使用。圆锥形施加器的半径和厚度分别从4 -7.5 mm和1到2 mm变化,以实现与不带圆锥体的HIFU换能器相似的声学分布。研究了与3D打印兼容的三种不同材料:聚乳酸,丙烯腈丁二烯苯乙烯和低密度聚乙烯。在COMSOL Multiphysics中实现了仿真

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号