Aluminium coated polysilicon (Al-polySi) nanoresonator has been used to design a urea biosensor. The etching characteristic of Al by the ammonium hydroxide from urea hydrolysis in the presence of urease was used to estimate the urea concentration. Urease immobilized on Al coated polySi indicated smoothening of the surface with increasing urea concentration, confirming the etching of Al. The nanoresonator could estimate between 10-1000 nM concentrations of urea. The sensing was not affected in the presence of interferents such as uric acid and ascorbic acid, thus making it specific for urea. The sensor was found to be highly sensitive and specific and can have potential application in the field of medical diagnostics and environmental monitoring.
机译:铝涂层多晶硅(Al-polySi)纳米谐振器已用于设计尿素生物传感器。在存在脲酶的情况下,由来自尿素水解的氢氧化铵对Al的蚀刻特性用于估计尿素浓度。固定在涂有Al的多晶硅上的脲酶表明,随着尿素浓度的增加,表面变得光滑,从而证实了Al的蚀刻。纳米谐振器可以估计10-1000 nM浓度的尿素。在存在诸如尿酸和抗坏血酸之类的干扰物时,传感不会受到影响,因此使其对尿素具有特异性。该传感器被发现具有很高的灵敏度和特异性,在医学诊断和环境监测领域具有潜在的应用前景。