In this study, we have demonstrated the usability of an epitaxial lead zirconate titanate (Epi-PZT) thin film for a piezoelectric micromachined ultrasonic transducer (pMUT) imager. At first, transmitting abilities and receiving sensitivities of the single-element pMUT based on Epi PZT and polycrystalline PZT (Poly-PZT) were evaluated as a fundamental study. Then, the 8-ch linear array system driven by a commercially-available pulser board was built, and successfully formed a focused ultrasonic beam. Finally, we obtained the ultrasonic images of a polydimethylsiloxane-based phantom by using the Epi-PZT and Poly-PZT pMUT linear array, and experimentally proved that Epi-PZT pMUT array can obtain higher contrast image compared to Poly-PZT device.
机译:在这项研究中,我们已经证明了外延钛酸锆钛酸铅(Epi-PZT)薄膜可用于压电微机械超声换能器(pMUT)成像器。首先,评估了基于Epi PZT和多晶PZT(Poly-PZT)的单元素pMUT的发射能力和接收灵敏度,作为基础研究。然后,构建了由市售脉冲发生器板驱动的8通道线性阵列系统,并成功地形成了聚焦超声束。最后,我们通过使用Epi-PZT和Poly-PZT pMUT线性阵列获得了基于聚二甲基硅氧烷的体模的超声图像,并通过实验证明Epi-PZT pMUT阵列与Poly-PZT器件相比可以获得更高的对比度图像。